Everyone is searching.: PCB Equipment 、 IC Substrate Equipment 、 JSW Vacuum Laminator 、 Vacuum Leveling Machine
Salt spray testing machine
1.Main specifications: 60L, 90L, 120L
2.Test types: NSS (neutral), ACSS (acidic), AASS (acetate spray test)
3.Salt water spray test: laboratory: 35 ± 1 ℃; Volatility ≤ ± 0.5 ℃; Uniformity ≤± 2 ℃
4.Corrosion resistance test: Laboratory: 50 ± 1 ℃; Volatility ≤ ± 0.5 ℃; Uniformity ≤± 2 ℃
5.Spray volume 1~2ml/h adjustable (ml/80cm2/h standard requires testing for 16 hours to calculate the average volume)
Operate and set according to JIS, ASTM, CNS, GB standard parameters
| A. | Salt water spray test: NSS, ACSS | |
| 1. | test room: | 35 ℃± 1 ℃ fluctuation: ≤± 0.5 ℃ uniformity: ≤ 2 ℃ |
| 2. | Saturated air tank: | 47℃ ± 1℃ |
| B. | Corrosion resistance test: CASS | |
| 1. | test room: | 50 ℃± 1 ℃ fluctuation: ≤± 0.5 ℃ uniformity: ≤ 2 ℃ |
| 2. | Saturated air tank: | 63℃ ± 1℃ |
| 1. | Materials used |
| a. | The test case body is made of light gray impact resistant PVC board with a thickness of 5mm and a durable temperature of 85 ℃. (Taiwan's South Asian card). The edge is surrounded by a three-dimensional reinforcement with a thickness of 10mm, and the structure is firm. |
| b. | The laboratory cover is sealed with PVC impact resistant board with a thickness of 5mm, which does not distort or deform during high temperature use. The inner and outer layers of the edge are thickened to prevent distortion and deformation caused by long-term high temperature. Thickness 15mm |
| c. | Hidden integrated reagent replenishment bottle, easy to clean and operate. |
| d. | The saturated air tank adopts SUS # 304 stainless steel high-pressure resistant tank for the best insulation effect (Taiwan Taizhan Electric Heating) |
| e. | The sample placement table is made of 8mm heat-resistant ABS board with strong load-bearing capacity. The test sample holder adopts a flat indexing type, which can adjust the angle arbitrarily. The fog drops evenly on all four sides, and the fog is completely consistent. The test results are accurate, and the number of test samples placed is large. (Can be customized according to customer product requirements) |
| 2. | spray system |
| a. | Adopting the Bernoulli principle to extract saltwater and then atomize it uniformly, without any obstruction or crystallization, ensuring the standard of continuous testing |
| b. | Nozzle: made of special glass nozzle, which can adjust the size of spray volume and spray angle |
| c. | The amount of spray can be adjusted from 1 to 2ml/h (ml/80cm2/h standard requires the average amount after 16 hours of testing). The measuring cylinder adopts embedded installation, with a beautiful and neat appearance, easy observation, and reduces instrument installation space |
| d. | Spray mode: continuous spray |
| 3. | Air supply system |
| a. | A section is roughly adjusted by 2Kg/cm2, and the imported air filter is equipped with drainage function |
| b. | The second stage involves precise adjustment of 1Kg/cm2 and 1/4 pressure gauge, displaying accurate and precise readings |
| 4. | heating system |
| a. | Adopting direct heating method, the heating speed is fast and the standby time is reduced. When the temperature reaches, it automatically switches to a constant temperature state, with precise temperature and low power consumption. Pure titanium heating tube, resistant to acid and alkali corrosion, with an ultra long service life |
| 5. | Control system |
| a. | The laboratory and pressure tank both use LCD type dual character temperature controllers with microcomputer automatic calculation function, with a control error of ± 1.0 ℃ (2 from Nippon Paint). The circuit board has undergone moisture-proof and anti-corrosion treatment, with high precision and long service life |
| b. | The laboratory heating tank adopts a liquid expansion safety temperature controller from 0 to 120 ℃ (1 EGO from Italy) |
| c. | LCD digital display time controller, power outage memory function 0.1S~999hr (1 OMRON from Japan) |
| d. | Accumulator 0-99999hr (1 VEMER from Italy) |
| e. | Relay (Japan OMRON Group 1) |
| f. | Swing board with light switch, capable of continuous operation 25000 times |
| 1. | When the water level is low, the power is automatically cut off and the safety warning light device lights up to display |
| 2. | When the temperature exceeds the limit, the heater power supply will be automatically cut off, and the safety warning light device will light up and display |
| 3. | When the water level of the test drug (saline) is low, the safety warning light device lights up and displays an alarm prompt |
| 4. | Leakage protection function, preventing personal injury and instrument failure caused by circuit leakage or short circuit |